|
Vinyl Cyclohexene Dioxide |
|
Method No.: |
PV2083 |
|
Matrix: |
Air |
|
Target concentration: |
10 ppm (57 mg/m³) (ACGIH TWA TLV) |
|
Procedure: |
Samples are collected by drawing a known volume of air through
an XAD-2 tube. Samples are desorbed with 1 mL of carbon disulfide
(CS2) for one hour and analyzed by gas chromatography
using a flame ionization detector (GC-FID). |
|
Recommended Air Volume and sampling rate: |
10 L at 0.2 L/min |
|
Reliable quantitation limit: |
0.54 ppm (3.095 mg/m3) |
|
Status of method: |
Partially Evaluated Method. This method has been subjected to
established evaluation procedures, and is presented for information
and trial use. |
|
Date: January, 1996 |
Chemist: Wayne Potter |
Organic Service Branch I OSHA Salt Lake Technical
Center Salt Lake City, UT 84165-1802
- Introduction
1.1 Background
1.1.1 History
The OSHA Technical Center has received several requests for
sampling and analytical information for vinyl cyclohexene dioxide
(VCD). The threshold limit value (TLV) for VCD is 10 ppm. The purpose
of this study was to obtain data on retention, storage and desorption
of VCD at the TLV level.
1.1.2 Toxic effects (This section is for information only and
should not be taken as the basis of OSHA policy.) (Ref. 5.1)
Vinyl cyclohexene dioxide is irritating to the eyes and skin, and
exposure can induce acute respiratory tract irritation and pulmonary
congestion. VCD is a direct-acting mutagen and it induces tumors both
local and distant from the site of application after topical treatment
or after parenteral injection. The rat LC50 was 800 ppm.
This atmosphere could not exist except in a state of supersaturation.
Given the reported vapor pressure of 0.1 torr at 20ºC the
corresponding saturated vapor concentration was 132 ppm in air. Even
at a temperature of 30ºC a
concentration of 800 ppm could not be reached. Hence, the risk of
acute intoxication by inhalation of VCD is considered slight.
1.1.3 Workplace exposure (Ref. 5.1)
Vinyl cyclohexene dioxide has been used by the plastics industry
since the 1950s in polymer formation and in other types of organic
syntheses. The National Institute of Occupational Safety and Health
(NIOSH) estimated that 6227 U.S. workers were exposed to VCD during
1981-1983.
1.1.4 Physical properties and other descriptive information (Ref.
5.1, 5.2)
Synonyms: |
7-Oxabicyclo(4.1.0)heptane,3-(epoxyethyl)-;
1,2-Epoxy-4-(epoxyethyl)
cyclohexane; 1-Epoxyethyl-3,4-epoxycyclohexane; Vinyl
cyclohexene diepoxide; 1-Vinyl-3-cyclohexane
dioxide; 4-Vinyl-1-cyclohexane diepoxide |
CAS
number: |
106-87-6 |
IMIS: |
2581 |
RTECS: |
RN8640000 |
Molecular
weight: |
140.18 |
Flash
point: |
110ºC
(230 ºF)(oc) |
Boiling
point: |
227ºC |
Freezing
point: |
-55ºC |
Color: |
clear, colorless
liquid or pale yellow liquid |
Specific
gravity: |
1.0986 at
20ºC |
Molecular
formula: |
C8H12O2
|
Structural
formula: |
|
The analyte air concentrations throughout this method are based
on the recommended sampling and analytical parameters. Air
concentrations listed in ppm are referenced to 25ºC and 101.3 kPa (760
mmHg).
1.2 Limit defining parameters
1.2.1 Detection limit of the overall procedure (DLOP)
The detection limit of the overall procedure is 9.28 µg per sample (0.162 ppm or 0.928 mg/m³).
This is the amount of analyte spiked on the sampler that will give a
response that is significantly different from the background response
of a sampler blank.
The DLOP is defined as the concentration of analyte that gives a
response (YDLOP) that is significantly different (three
standard deviations (SDBR)) from the background response
(YBR).
YDLOP - YBR =
3(SDBR)
The direct measurement of YBR and SDBR in
chromatographic methods is typically inconvenient, and difficult
because YBR is usually extremely low. Estimates of these
parameters can be made with data obtained from the analysis of a
series of samples whose responses are in the vicinity of the
background response. The regression curve obtained for a plot of
instrument response versus concentration of analyte will usually be
linear. Assuming SDBR and the precision of data about the
curve are similar, the standard error of estimate (SEE) for the
regression curve can be substituted for SDBR in the above
equation. The following calculations derive a formula for the
DLOP:
Yobs = observed response
Yest = estimated response from
regression curve n = total no. of data
points k = 2 for a linear regression
curve
At point YDLOP on the regression curve
YDLOP = A(DLOP) +
YBR
A = analytical sensitivity
(slope)
therefore
Substituting 3(SEE) + YBR for YDLOP gives
The DLOP is measured as mass per sample and expressed as equivalent
air concentrations, based on the recommended sampling parameters. Ten
samplers were spiked with equal descending increments of analyte, such
that the lowest sampler loading was 22.80 µg/sample. This is the amount, when spiked on
a sampler, that would produce a peak approximately 10 times the
background response for the sample blank. These spiked samplers, and
the sample blank were analyzed with the recommended analytical
parameters, and the data obtained used to calculate the required
parameters (A and SEE) for the calculation of the DLOP. Values of 9.44
and 29.214 were obtained for A and SEE respectively. DLOP was
calculated to be 9.28 µg/sample (0.162
ppm or 0.928 mg/m3).
Table 1.2.1 Detection Limit of the Overall
Procedure
|
mass per sample (µg) |
area counts (µV-s) |
|
0 22.8 28.5 34.2 39.9 45.6 51.3 57.0 62.7 68.4 74.1 |
0 211 241 264 383 446 459 504 622 625 658 |
|
Figure 1.2.1 Plot of data to determine the
DLOP/RQL
1.2.2 Reliable quantitation limit (RQL)
The reliable quantitation limit is 30.95 µg per sample (0.54
ppm or 3.095 mg/m3). This is the amount of analyte spiked
on a sampler that will give a signal that is considered the lower
limit for precise quantitative measurements.
Figure 1.2.2 Chromatogram of the RQL
The RQL is considered the lower limit for precise quantitative
measurements. It is determined from the regression line data obtained
for the calculation of the DLOP (Section 1.2.1), providing at least
75% of the analyte is recovered. The RQL is defined as the
concentration of analyte that gives a response (YRQL) such
that
YRQL - YBR =
10(SDBR)
therefore:
RQL= 30.95µg per
sample (0.54 ppm or 3.095 mg/m3)
- Sampling Procedure
2.1 Apparatus
2.1.1 Samples are collected using a personal sampling pump
calibrated, with the sampling device attached, to within ±5% of the
recommended flow rate.
2.1.2 Samples are collected with tubes 7 cm × 4 mm i.d. × 6 mm o.d.
glass sampling tubes packed with two sections of XAD-2. The front
section contains 80 mg and the back section contains 40 mg of XAD-2.
The sections are held in place with glass wool plugs and are separated
by a glass wool plug. For this evaluation, commercially prepared
sampling tubes were purchased from SKC Inc., (Eighty Four, PA) catalog
No. 226-30, Lot 879.
2.2 Technique
2.2.1 Immediately before sampling, break off the ends of the
sampling tube. All tubes should be from the same lot.
2.2.2 Attach the sampling tube to the pump with flexible tubing. It
is desirable to utilize sampling tube holders which have a protective
cover to shield the employee from the sharp, jagged end of the
sampling tube. Position the tube so that sampled air passes through
the front section of the tube first.
2.2.3 Air being sampled should not pass through any hose or tubing
before entering the sampling tube.
2.2.4 Attach the sampling tube vertically with the front section
pointing downward, in the worker's breathing zone, and positioned so
it does not impede work performance or safety.
2.2.5 After sampling for the appropriate time, remove the sample
and seal the tube with plastic end caps. Wrap each sample end-to-end
with a Form OSHA-21 seal.
2.2.6 Submit at least one blank sample with each set of samples.
Handle the blank sample in the same manner as the other samples except
draw no air through it.
2.2.7 Record sample volumes (in liters of air) for each sample,
along with any potential interferences.
2.2.8 Ship any bulk samples separate from the air samples.
2.2.9 Submit the samples to the laboratory for analysis as soon as
possible after sampling. If delay is unavoidable, store the samples in
a refrigerator.
2.3 Desorption efficiency
The desorption efficiencies of VCD were determined by liquid-spiking
the XAD-2 tubes with the analyte at 0.1 to 2 times the target
concentration. The loadings on the tubes were 57, 285 ,570 and 1140
µg of VCD. These samples were stored overnight at ambient
temperature and then desorbed with 1 mL of CS2 with 0.25
µL/mL p-cymene internal standard,
and analyzed by GC-FID. The average desorption efficiency over the
studied range was 97.0%.
Table 2.3 Desorption Efficiency of VCD
|
% Recovered |
|
0.1 × |
0.5 × |
1.0 × |
2.0 × |
Tube # |
57µg |
285µg |
570µg |
1140µg |
|
1 2 3 4 5 6 |
91.5 90.9 93.1 93.5 96.6 90.7 |
98.2 99.7 99.4 100.2 99.3 100.1 |
97.8 96.5 98.4 99.5 100.5 101.0 |
96.9 96.4 97.3 97.5 96.0 96.9 |
average |
92.7 |
99.5 |
99.0 |
96.8 |
overall average |
97.0 |
|
|
|
standard deviation |
±2.98 |
|
|
|
|
2.4 Retention efficiency
The glass wool in front of the front section of the XAD-2 tube was
pulled towards the end, so that none of it was in contact with the
XAD-2. The glass wool was spiked with 1140 µg VCD, and the XAD-2 tube had 10 L humid air
(80% RH at 25ºC) pulled through it at
0.2 L/min. The glass wool was spiked to determine if VCD would volatize
off the glass wool and collect onto the XAD-2. They were opened,
desorbed, and analyzed by GC-FID. The retention efficiency averaged
97.7%. The values in Table 2.4 were not corrected for desorption
efficiency. There was no VCD found on the glass wool indicating that all
of it vaporized off. There was no VCD on the back sections of the tubes,
indicating that no breakthrough occurred.
Table 2.4 Retention Efficiency of VCD
|
Tube # |
% Recovered |
|
Glass wool |
Front section |
Back section |
Total |
|
1 2 3 4 5 6 |
0.0 0.0 0.0 0.0 0.0 0.0 |
96.9 97.1 95.8 98.6 99.9 98.1 |
0 0 0 0 0 0 |
96.9 97.1 95.8 98.6 99.9 98.1 |
|
|
|
average |
97.7 |
|
2.5 Sample storage
The front sections of twelve sampling tubes were each spiked with
1140 µg (19.9 ppm) of VCD, and had 10
liters of dry air drawn through them. Six of the tubes were stored in
the refrigerator (-10ºC), and six
were stored at room temperature (25ºC). Twelve more tubes were spiked with
1140 µg of VCD, and had 10 liters of humid air (80% RH at 25ºC) drawn through them. Six tubes were
stored in the refrigerator (-10ºC),
and six were stored at room temperature (25ºC). Three of each type of samples were
analyzed after 7 days and the remaining three samples of each type after
14 days. The amounts recovered indicate that humidity and temperature
had no affect on the ability of XAD-2 to retain VCD over the 14 days
studied. The results are not corrected for desorption
efficiency.
Table 2.5 Storage Test for VCD
|
Time (days) |
%Recovery Humid Ambient |
%Recovery Humid Refrigerated |
%Recovery Dry Ambient |
%Recovery Dry Refrigerated |
|
average 14 14 14 |
99.8 101.0 100.1 100.3 101.8 100.9 97.4 |
97.8 100.1 102.9 100.3 99.3 97.0 98.7 |
98.2 99.9 99.3 99.1 100.8 99.8 100.3 |
100.6 98.0 98.9 99.2 101.3 95.0 100.2 |
average |
100.0 |
98.3 |
100.3 |
98.8 |
|
2.6 Recommended air volume and sampling rate.
Based on the data collected in this evaluation, 10 L air samples
should be collected at a sampling rate of 0.2 L/min.
2.7 Interferences (sampling)
2.7.1 It is not known if any compounds will severely interfere with
the collection of VCD on the sampling tubes. In general, the presence
of other contaminant vapors in the air will reduce the capacity of the
XAD-2 tube to collect VCD.
2.7.2 Suspected interferences should be reported to the laboratory
with submitted samples.
2.8 Safety precautions (sampling)
2.8.1 Attach the sampling equipment to the worker in such a manner
that it will not interfere with work performance or safety.
2.8.2 Follow all safety practices that apply to the work area being
sampled.
2.8.3 Wear eye protection when breaking the ends of the glass
sampling tubes.
- Analytical Procedure
3.1 Apparatus
3.1.1 The instrument used in this study was a gas chromatograph
equipped with a flame ionization detector, specifically a Hewlett
Packard model 5890.
3.1.2 A GC column capable of separating the analyte from any
interferences. The column used in this study was a 60 meter capillary
column with a 1.5 µm coating of
Rtx-Volatiles, with an I.D. of 0.32 mm.
3.1.3 An electronic integrator or some suitable method of measuring
peak areas.
3.1.4 Two milliliter vials with Teflon™-lined caps.
3.1.5 A 10 µL syringe or other
convenient size for sample injection.
3.1.6 Pipets for dispensing the desorbing solution. A Repipet®
dispenser was used in this study.
3.1.7 Volumetric flasks - 5 or 10 mL and other convenient sizes for
preparing standards.
3.2 Reagents
3.2.1 GC grade nitrogen, hydrogen, and air.
3.2.2 Vinyl cyclohexene dioxide (VCD), Reagent grade
3.2.3 Carbon disulfide (CS2), Reagent grade
3.2.4 p-Cymene (internal standard), Reagent grade
3.2.5 Desorbing solution was carbon disulfide with 0.25 µL/mL p-cymene internal
standard.
3.3 Standard preparation
3.3.1 At least two separate stock standards are prepared by
diluting a known quantity of VCD with the desorbing solution of carbon
disulfide with 0.25 µL/mL
p-cymene internal standard. The concentration of these stock
standards was 0.5 µL/mL or 549.3 µg/mL.
3.3.2 A third standard at a higher concentration was prepared to
check the linearity of the calibration. For this study, two analytical
standards were prepared at a concentration of 0.5 µL/mL (549.3 µg/mL), and one at 2.0 µL/mL (2197.2 µg/mL) VCD in the desorbing
solution.
3.4 Sample preparation
3.4.1 Sample tubes are opened and the front and back section of
each tube are placed in separate 2 mL vials.
3.4.2 Each section is desorbed with 1 mL of the desorbing solution
of carbon disulfide with 0.25 µL/mL
p- cymene internal standard.
3.4.3 The vials are sealed immediately and allowed to desorb for 60
minutes with occasional shaking.
3.5 Analysis
3.5.1 Gas chromatograph conditions.
Chromatogram: Injection
size: |
1 µL |
Flow rates (mL/min)
|
|
Nitrogen (make-up): |
30 |
Hydrogen(carrier): |
2 |
Hydrogen(detector): |
40 |
Air: |
420 |
Temperatures
(ºC) |
|
Injector: |
200 |
Detector: |
220 |
Column: |
150º for 5 min then
10º/min to 200º for 10
min. |
Figure 3.5.1 Chromatogram of an analytical standard at the target
concentration. Peak identification: (1) carbon disulfide, (2)
p-cymene, (3) VCD isomers.
3.5.2 Peak areas are measured by an integrator or other suitable
means.
3.6 Interferences (analytical)
3.6.1 Any compound that produces a response and has a similar
retention time as the analyte is a potential interference. If any
potential interferences were reported, they should be considered
before samples are desorbed. Generally, chromatographic conditions can
be altered to separate an interference from the analyte.
Figure 3.6.1 Mass spectra of the isomers of VCD.
3.6.2 When necessary, the identity or purity of an analyte peak may
be confirmed by GC-mass spectrometer or by another analytical
procedure.
3.7 Calculations
3.7.1 The instrument was calibrated with a standard of 570 µg/mL VCD in the desorbing solution. The
linearity of the calibration was checked with a standard of 2160µg/mL.
3.7.2 If the calibration is non-linear, two or more standard at
different concentrations must be analyzed, bracketing the samples, so
a calibration curve can be plotted and sample values obtained.
3.7.3 To calculate the concentration of analyte in the air sample
the following formulas are used:
mass of analyte in sample = (mg/ml)(desorption
volumn)
desorption efficiency
number of moles of analyte = mass of analyte in
sample
molecular weight
Volume the analyte will occupy at 25ºC and 760 mmHg is number of moles of
analyte times the molar volume at 25ºC and 760 mmHg.
ppm = (volume analyte
occupies)(106)
molecular weight
3.7.4 The above equations can be consolidated to the following
formula.
µg/mL = concentration of analyte
in sample or standard 24.46 = molar volume (liters/mole) at
25ºC and 760 mmHg MW =
molecular weight (g/mole) DV = desorption volume 10 L = 10
liter air sample DE = desorption efficiency * All units must
cancel.
3.7.5 This calculation is done for each section of the sampling
tube and the results added together.
3.8 Safety precautions (analytical)
3.8.1 Avoid skin contact and inhalation of all chemicals.
3.8.2 Wear safety glasses, gloves and a lab coat at all times while
in the laboratory areas.
- Recommendations for Further Study
Collection studies need to be performed from a dynamically generated
test atmosphere.
- References
5.1 "Documentation of the Threshold Limit Values and Biological
Exposure Indices", Fifth Edition, American Conference of Governmental
Industrial Hygienists Inc., Cincinnati, OH, 1986, pp. 1708-1710.
5.2 Sweet, D.V., "Registry of Toxic Effects of Chemical Substances",
1987, Vol. 3, U.S.D.H.E.W., Index Number RN8640000, p.
3058.
|